Ultra-Technekow by is a Prescription medication manufactured, distributed, or labeled by Curium US LLC. Drug facts, warnings, and ingredients follow.
The Ultra-Technekow™ V4 Generator is prepared with fission-produced molybdenum Mo-99 adsorbed onto alumina in a column shielded by lead, tungsten, or depleted uranium. The column assembly and shielding are encased in a plastic container that is covered with a plastic elution hood. The elution hood has an opening for the column assembly double inlet needles and an opening for the single outlet needle. The needles accommodate the sterile eluant vials and sterile evacuated collection vials. A sterile vial containing a bacteriostat is supplied with the generator for the customer to aseptically seal the outlet needle after each elution.
This terminally sterilized generator provides a closed system for the production of sterile metastable technetium Tc-99m, which is produced by the decay of molybdenum Mo-99. Incorporated between the column outlet and the collection vial is a sterile 0.22 micrometer filter. Sterile, non-pyrogenic isotonic solutions of Sodium Pertechnetate Tc 99m can be obtained conveniently by periodic aseptic elution of the generator. These solutions should be clear, colorless, and free from any particulate matter. The Sodium Pertechnetate
Tc 99m Injection is suitable for intravenous injection and direct instillation.
The carrier-free solution may be used as is, or diluted to the proper concentration. Over the life of the generator, an elution will contain an amount of technetium Tc-99m in direct proportion to the quantity of Mo-99 decay since the previous elution of the generator. The quantity of Tc-99m in the eluate is determined by quantity of Mo-99 on the column, and the elapsed time between elutions.
Each eluate of the generator should not contain more than the USP limit of 0.15 kilobecquerel molybdenum Mo-99 per megabecquerel technetium Tc-99m (0.15 microcurie Mo-99 per millicurie Tc-99m) per administered dose at the time of administration and an aluminum ion concentration of not more than 10 micrograms per milliliter of the generator eluate, both of which must be determined by the user before administration.
Since the eluate does not contain an antimicrobial agent, it should not be used after 12 hours from the time of generator elution.
Technetium Tc-99m decays by isomeric transition with a physical half-life of 6 hours. The principal photon that is useful for detection and imaging studies is listed in Table 1.
Table 1. Principal Radiation Emission Data
Radiation |
Mean Percent |
Energy |
Gamma-2 |
89.07 |
140.5 |
The specific gamma ray constant for technetium Tc-99m is 0.795 R/hr-mCi at 1 cm. The first half-value layer is 0.023 cm of lead (Pb). A range of values for the relative attenuation of the radiation emitted by this radionuclide that results from interposition of various thicknesses of Pb is shown in Table 2. For example, the use of 0.27 cm thickness of Pb will attenuate the radiation emitted by a factor of about 1000.
Table 2. Radiation Attenuation by Lead Shielding
Shield |
Coefficient |
0.023 |
0.5 |
0.09 |
10-1 |
0.18 |
10-2 |
0.27 |
10-3 |
Molybdenum Mo-99 decays to technetium Tc-99m with a molybdenum Mo-99 half-life of 2.75 days, or 66 hours (see Table 3). The physical decay characteristics of molybdenum Mo-99 are such that only 88.6% of the decaying molybdenum Mo-99 atoms form technetium Tc-99m. Generator elutions may be made at any time, but the amount of technetium Tc-99m available will depend on the interval measured from the last elution. Approximately 47% of the maximum available technetium Tc-99m is reached after 6 hours and 95% after 23 hours. To correct for physical decay of molybdenum Mo-99 and technetium Tc-99m, the fractions that remain at selected intervals of time are shown in Tables 3 and 4.
Table 3. Physical Decay Chart; Molybdenum Mo-99, Half-Life 66 Hours
Days |
Percent Remaining |
Days |
Percent Remaining |
0 |
100 |
10 |
8 |
1 |
78 |
11 |
6 |
2 |
60 |
12 |
5 |
3 |
47 |
13 |
4 |
4 |
37 |
14 |
3 |
5 |
28 |
15 |
2 |
6 |
22 |
20 |
0.6 |
7 |
17 |
25 |
0.2 |
8 |
13 |
30 |
0.05 |
9 |
10 |
Table 4. Physical Decay Chart; Technetium Tc-99m, Half-Life 6 Hours
Hours |
Percent |
Hours |
Percent |
0* |
100 |
9 |
35 |
1 |
89 |
10 |
32 |
2 |
79 |
11 |
28 |
3 |
71 |
12 |
25 |
4 |
63 |
14 |
20 |
5 |
56 |
16 |
16 |
6 |
50 |
18 |
13 |
7 |
45 |
24 |
6 |
8 |
40 |
*Calibration Time
The pertechnetate ion distributes in the body similarly to the iodide ion but is not organified when trapped in the thyroid gland. Pertechnetate concentrates in the thyroid gland, salivary glands, stomach and choroid plexus. After intravenous administration it gradually equilibrates with the extracellular space. A fraction is promptly excreted via the kidneys.
Following the administration of Sodium Pertechnetate Tc 99m as an eye drop, the drug mixes with tears within the conjunctival space. Within seconds to minutes it leaves the conjunctival space and escapes into the inferior meatus of the nose through the nasolacrimal drainage system. During this process the pertechnetate ion passes through the canaliculi, the lacrimal sac and the nasolacrimal duct. In the event of any anatomical or functional blockage of the drainage system there will be a backflow resulting in tearing (epiphora). Thus the pertechnetate escapes the conjunctival space in the tears.
While the major part of the pertechnetate escapes within a few minutes of normal drainage and tearing, it has been documented that there is some degree of transconjunctival absorption with turnover of 1.5% per minute in normal individuals, 2.1% per minute in patients without any sac and 2.7% per minute in patients with inflamed conjunctiva due to chronic dacryocystitis. Individual values may vary but these rates are probably representative and indicate that the maximum possible pertechnetate absorbed will remain below one thousandth of that used in other routine diagnostic procedures.
The Ultra-Technekow™ V4 generator is a source of sodium pertechnetate Tc 99m for use in the preparation of FDA-approved diagnostic radiopharmaceuticals, as described in the labeling of these diagnostic radiopharmaceutical kits.
Sodium Pertechnetate Tc 99m is used IN ADULTS as an agent for:
Thyroid Imaging
Salivary Gland Imaging
Urinary Bladder Imaging (direct isotopic cystography) for detection of vesico-ureteral reflux
Nasolacrimal Drainage System Imaging (dacryoscintigraphy)
Sodium Pertechnetate Tc 99m is used IN PEDIATRIC PATIENTS as an agent for:
Thyroid Imaging
Urinary Bladder Imaging (direct isotopic cystography) for the detection of vesico-ureteral reflux
Radiation risks associated with the use of Sodium Pertechnetate Tc 99m are greater in pediatric patients than in adults and, in general, the younger the patient the greater the risk owing to greater absorbed radiation doses and longer life expectancy. These greater risks should be taken firmly into account in all benefit risk assessments involving pediatric patients.
Long-term cumulative radiation exposure may be associated with an increased risk of cancer.
Only use generator eluant specified for use with the Ultra-Technekow™ V4 Generator. Do not use any other generator eluant or saline from any other source.
Radiopharmaceuticals should be used only by physicians who are qualified by training and experience in the safe use and handling of radionuclides and whose experience and training have been approved by the appropriate government agency authorized to license the use of radionuclides.
As in the use of any radioactive material, care should be taken to minimize radiation exposure to the patient consistent with proper patient management and to ensure minimum radiation exposure to occupational workers.
After the termination of the nasolacrimal imaging procedure, blowing the nose and washing the eyes with sterile distilled water or an isotonic sodium chloride solution will further minimize the radiation dose.
Since the eluate does not contain an antimicrobial agent, it should not be used after 12 hours from time of generator elution.
No long-term animal studies have been performed to evaluate carcinogenic or mutagenic potential or whether Sodium Pertechnetate Tc 99m may affect fertility in males or females.
In animal reproductive studies, Sodium Pertechnetate Tc 99m (as free pertechnetate) has been shown to cross the placental barrier. It is not known whether Sodium Pertechnetate Tc 99m can cause fetal harm when administered to a pregnant woman or can affect reproductive capacity.
Sodium Pertechnetate Tc 99m should be given to pregnant women only if the expected benefits to be gained clearly outweigh the potential hazards.
Ideally, examinations using radiopharmaceutical drug products - especially those elective in nature - of women of childbearing capability should be performed during the first ten days following the onset of menses.
Technetium Tc-99m is excreted in human milk during lactation, therefore, formula-feedings should be substituted for breast-feedings.
See INDICATIONS AND USAGE and DOSAGE AND ADMINISTRATION sections. Also see the description of additional risk under WARNINGS.
Sodium Pertechnetate Tc 99m is administered by intravenous injection. When imaging the nasolacrimal drainage system, instill the Sodium Pertechnetate Tc 99m by the use of a device such as a micropipette or similar method which will ensure the accuracy of the dose.
For imaging the urinary bladder and ureters (direct isotopic cystography), the Sodium Pertechnetate Tc 99m is administered by direct instillation aseptically into the bladder via a urethral catheter, following which the catheter is flushed with approximately 200 mL of sterile saline directly into the bladder.
The suggested dose ranges employed for various diagnostic indications in the average ADULT PATIENT (70 kg) are as follows:
Vesico-ureteral imaging: 18.5 to 37 MBq (0.5 to 1 mCi)
Thyroid gland imaging: 37 to 370 MBq (1 to 10 mCi)
Salivary gland imaging: 37 to 185 MBq (1 to 5 mCi)
Nasolacrimal drainage system: Maximum dose of 3.7 MBq (100 µCi)
The recommended dosages in PEDIATRIC PATIENTS are:
Vesico-ureteral imaging: 18.5 to 37 MBq (0.5 to 1 mCi)
Thyroid gland imaging: 2.22 to 2.96 MBq (60 to 80 µCi) per kg body weight
The patient dose should be measured by a suitable radioactivity calibration system immediately prior to administration.
Parenteral drug products should be inspected visually for particulate matter and discoloration prior to administration whenever solution and container permit. If the solution is discolored, discontinue use of the generator immediately. The solution to be administered as the patient dose should be clear, colorless, and contain no particulate matter.
The estimated absorbed radiation doses to an average ADULT and PEDIATRIC patient from an intravenous injection of various doses of Sodium Pertechnetate Tc 99m distributed uniformly in the total body are shown in Tables 5 and 6.
Table 5. Absorbed Radiation Doses from Intravenous Injection
Organ |
Absorbed Radiation Dose (mGy) for a 1110 MBq (30mCi) dose |
Adrenals |
4.1 |
Urinary Bladder Wall |
20 |
Bone Surfaces |
6.2 |
Brain |
2.2 |
Breasts |
2 |
Gallbladder Wall |
8.3 |
Stomach Wall |
29 |
Small Intestine |
18 |
ULI Wall |
63 |
LLI Wall |
23 |
Heart Wall |
3.5 |
Kidneys |
6 |
Liver |
4.7 |
Lungs |
2.9 |
Muscle |
3.6 |
Ovaries |
11 |
Pancreas |
6.3 |
Red Marrow |
4.1 |
Skin |
2 |
Spleen |
4.8 |
Testes |
3.1 |
Thymus |
2.7 |
Thyroid |
24 |
Uterus |
9 |
Remaining Tissues |
3.9 |
Effective Dose (mSv) |
14 |
To obtain radiation absorbed dose in rads (30 mCi dose) from the above table, divide individual organ values by a factor of 10 (does not apply for effective dose).
Table 6. Pediatric Absorbed Radiation Doses (mGy) from Intravenous Injection
Age |
15 years |
10 years |
5 years |
1 year |
Administered activity in MBq (mCi) |
1110 |
740 |
555 |
370 |
Organ | ||||
Adrenals |
5.3 |
5.4 |
6.2 |
7.1 |
Urinary Bladder Wall |
26 |
22 |
18 |
22 |
Bone Surfaces |
7.6 |
7.5 |
8.1 |
10 |
Brain |
2.8 |
3.1 |
3.7 |
4.5 |
Breasts |
2.6 |
2.6 |
3.2 |
4.1 |
Gallbladder Wall |
11 |
12 |
13 |
13 |
Stomach Wall |
38 |
36 |
43 |
59 |
Small Intestine |
22 |
23 |
26 |
30 |
ULI Wall |
81 |
89 |
110 |
140 |
LLI Wall |
31 |
33 |
40 |
48 |
Heart Wall |
4.5 |
4.6 |
5.2 |
6.4 |
Kidneys |
7.2 |
6.9 |
7.8 |
8.5 |
Liver |
6 |
6.7 |
8 |
9.1 |
Lungs |
3.8 |
3.8 |
4.4 |
5.3 |
Muscle |
4.5 |
4.5 |
5 |
6 |
Ovaries |
14 |
13 |
14 |
17 |
Pancreas |
8.1 |
8.2 |
8.9 |
10 |
Red Marrow |
5.1 |
5 |
5.2 |
6 |
Skin |
2.5 |
2.6 |
3.2 |
3.8 |
Spleen |
6 |
6 |
6.7 |
7.8 |
Testes |
4.1 |
4.3 |
4.9 |
6 |
Thymus |
3.6 |
3.5 |
4.2 |
5.3 |
Thyroid |
40 |
41 |
67 |
81 |
Uterus |
11 |
11 |
12 |
14 |
Remaining Tissues |
4.8 |
4.8 |
5.4 |
6.4 |
Effective Dose (mSv) |
19 |
19 |
23 |
29 |
To obtain radiation absorbed dose in rads (30 mCi dose) from the above table, divide individual organ values by a factor of 10 (does not apply for effective dose).
The estimated absorbed radiation doses to an ADULT patient from the nasolacrimal imaging procedure using a maximum dose of 3.7 megabecquerels (100 microcuries) of Sodium Pertechnetate Tc 99m are shown in Table 7.
Table 7. Absorbed Radiation Doses from Dacryoscintigraphy
Tissue |
3.7 MBq (100 µCi) |
|
mGy |
rad |
|
Eye Lens: If lacrimal fluid turnover is 16%/min |
0.140 |
0.014 |
If lacrimal fluid |
0.022 |
0.002 |
If drainage |
4.020 |
0.402 |
Total Body* |
0.011 |
0.001 |
Ovaries* |
0.030 |
0.003 |
Testes* |
0.009 |
0.001 |
Thyroid* |
0.130 |
0.013 |
*Assuming no blockage of draining system.
In pediatric patients, an average 30 minute exposure to 37 MBq (1 mCi) of Tc-99m pertechnetate following instillation for direct cystography, will result in the following estimated radiation doses:
Table 8. Absorbed Radiation Doses from Cystography (PEDIATRIC)
Age |
Bladder wall dose, mGy (rad) |
Gonadal dose, |
1 year |
3.6 (0.36) |
0.15 (0.015) |
5 years |
2.0 (0.2) |
0.095 (0.0095) |
10 years |
1.3 (0.13) |
0.066 (0.0066) |
15 years |
0.92 (0.092) |
0.046 (0.0046) |
The Ultra-Technekow™ V4 (Technetium Tc 99m) Generators contain the following amount of molybdenum Mo-99 at the date and time of calibration stated on the label.
Catalog No.
9010 37 gigabecquerels (1.0 curie)
NDC: 69945-010-03
9015 55.5 gigabecquerels (1.5 curies)
NDC: 69945-015-04
9020 74 gigabecquerels (2.0 curies)
NDC: 69945-020-05
9025 92.5 gigabecquerels (2.5 curies)
NDC: 69945-025-06
9030 111 gigabecquerels (3.0 curies)
NDC: 69945-030-07
9035 129.5 gigabecquerels (3.5 curies)
NDC: 69945-035-08
9051 185 gigabecquerels (5.0 curies)
NDC: 69945-051-09
9060 222 gigabecquerels (6.0 curies)
NDC: 69945-060-10
9075 277.5 gigabecquerels (7.5 curies)
NDC: 69945-075-11
9110 407 gigabecquerels (11.0 curies)
NDC: 69945-110-12
9140 518 gigabecquerels (14.0 curies)
NDC: 69945-140-13
9160 592 gigabecquerels (16.0 curies)
NDC: 69945-160-14
9190 703 gigabecquerels (19.0 curies)
NDC: 69945-190-15
Each generator is supplied with the following components for the elution of the generator:
1 - Technestat™ Vial, 5 mL, containing 0.5 mL of 1.5 mg/mL methylparaben and 0.2 mg/mL propylparaben, sterile, non-pyrogenic
1 - Package Insert
SUPPLIED SEPARATELY
30 - Evacuated Collecting Vials, 30 mL, sterile, non-pyrogenic, supplied with:
90 - Radioactive Materials Labels – Collection Vial (30 en, 30 fr, 30 es)
90 - Radioactive Materials Labels – Elution Shield (30 en, 30 fr, 30 es)
1 - Package Insert
30 - Generator Eluant, 0.9% sodium chloride, sterile, non-pyrogenic, available in 5,
10 mL, or 20 mL volumes, with 1 package insert. The eluant does not contain an antimicrobial agent.
Storage
Store generator and Sodium Pertechnetate Tc 99m solution at controlled room temperature 20-25°C (68-77°F) [see USP Controlled Room Temperature].
Expiration Date
The generator should not be used after the expiration date stated on the label.
The expiration time of the Sodium Pertechnetate Tc 99m solution is not later than 12 hours after time of elution. If the eluate is used to reconstitute a kit, the radiolabeled kit should not be used after 12 hours from the time of generator elution or after the expiration time stated on the labeling for the prepared drug, whichever is earlier.
NOTE 1: Immediately upon delivery, the generator should be placed within a minimum of one-inch of lead shielding in such a manner
so as to minimize radiation exposure to attending personnel.
NOTE 2: Wear waterproof gloves during the elution procedure and during subsequent reconstitution of kits with the eluate.
NOTE 3: Use a shielded syringe to withdraw patient dose or to transfer Sodium Pertechnetate Tc 99m into mixing vials during kit
reconstitution.
NOTE 4: The needles in the generator are sterile beneath their covers, and the generator has been cleaned underneath the top cover.
Additional disinfection of these areas with agents containing alcohol may unfavorably influence the Tc-99m yield.
Eluting the generator every 24 hours will provide optimal amounts of Sodium Pertechnetate Tc 99m. However, the generator may be eluted whenever sufficient amounts of technetium Tc-99m have accumulated within the column.
For Example
Time After |
Approximate Yield |
1 |
10 |
2 |
19 |
3 |
27 |
4 |
35 |
5 |
41 |
6 |
47 |
Elution
Subsequent Elutions
Vacuum Loss
If the vacuum in the collecting vial is lost, do not attempt to re-evacuate the vial, but discard and use a new collecting vial.
Expired Generator Disposal
This generator may be received, used and administered only by authorized persons in designated clinical settings. Its receipt, storage, use, transfer and disposal are subject to the regulations and/or appropriate licenses of local competent official organizations.
Curium and the Curium logo are trademarks of a Curium company.
©2018 Curium US LLC. All Rights Reserved.
Manufactured by:
Curium US LLC
Maryland Heights, MO 63043 USA
Made in USA
R12/2018
STERILE
9010-9190
CURIUM™
Ultra-Technekow™ V4
(TECHNETIUM Tc 99m GENERATOR)
Parent Molybdenum-99 prepared from Fission Produced Molybdenum-99
Store at Controlled Room Temperature 20-25ºC (68-77ºF)
WARNING: EACH UTK-V4 GENERATOR ELUATE MUST BE TESTED FOR MOLYBDENUM-99 CONTENT BEFORE USE.
Eluate must be used within 12 hours of elution.
For the production of sterile, non-pyrogenic Sodium Pertechnetate Tc 99m injection.
Rx only
See package insert for directions for use.
Tc-99m Half Life = 6 hr
CAUTION
RADIOACTIVE MATERIAL
Patent/Brevet/Patente:
www.curiumpharma.com/us-patents
Manufactured by/Fabrique par/Fabricado por:
Curium US LLC
Maryland Heights, MO 63043
Made in USA/Fabrique aux Etats-Unis/Hecho en E.U.A.
Distributed in Canada by/
Distribue au Canda par:
Curium Canada Inc.
Laval, QC, Canada, H7T-2R3
DIN: 02441608
CURIUM TM/MC
A901C0
R03/2019
Label
Ultra-Technekow™ V4
(TECHNETIUM Tc 99m GENERATOR)
Parent Molybdenum-99 prepared from Fission Produced Molybdenum-99
Store at Controlled Room Temperature 20-25ºC (68-77ºF)
WARNING: EACH UTK-V4 GENERATOR ELUATE MUST BE TESTED FOR MOLYBDENUM-99 CONTENT BEFORE USE.
Eluate must be used within 12 hours of elution.
For the production of sterile, non-pyrogenic Sodium Pertechnetate Tc 99m injection.
Rx only
See package insert for directions for use.
Tc-99m Half Life = 6 hr
CAUTION
RADIOACTIVE MATERIAL
Patent/Brevet/Patente:
www.curiumpharma.com/us-patents
Manufactured by/Fabrique par/Fabricado por:
Curium US LLC
Maryland Heights, MO 63043
Made in USA/Fabrique aux Etats-Unis/Hecho en E.U.A.
Distributed in Canada by/
Distribue au Canda par:
Curium Canada Inc.
Laval, QC, Canada, H7T-2R3
DIN: 02441608
CURIUM TM/MC
A901C0
R03/2019
Label
Ultra-Technekow™ V4
(TECHNETIUM Tc 99m GENERATOR)
Parent Molybdenum-99 prepared from Fission Produced Molybdenum-99
Store at Controlled Room Temperature 20-25ºC (68-77ºF)
WARNING: EACH UTK-V4 GENERATOR ELUATE MUST BE TESTED FOR MOLYBDENUM-99 CONTENT BEFORE USE.
Eluate must be used within 12 hours of elution.
For the production of sterile, non-pyrogenic Sodium Pertechnetate Tc 99m injection.
Rx only
See package insert for directions for use.
Tc-99m Half Life = 6 hr
CAUTION
RADIOACTIVE MATERIAL
Patent/Brevet/Patente:
www.curiumpharma.com/us-patents
Manufactured by/Fabrique par/Fabricado por:
Curium US LLC
Maryland Heights, MO 63043
Made in USA/Fabrique aux Etats-Unis/Hecho en E.U.A.
Distributed in Canada by/
Distribue au Canda par:
Curium Canada Inc.
Laval, QC, Canada, H7T-2R3
DIN: 02441608
CURIUM TM/MC
A901C0
R03/2019
Label
Ultra-Technekow™ V4
(TECHNETIUM Tc 99m GENERATOR)
Parent Molybdenum-99 prepared from Fission Produced Molybdenum-99
Store at Controlled Room Temperature 20-25ºC (68-77ºF)
WARNING: EACH UTK-V4 GENERATOR ELUATE MUST BE TESTED FOR MOLYBDENUM-99 CONTENT BEFORE USE.
Eluate must be used within 12 hours of elution.
For the production of sterile, non-pyrogenic Sodium Pertechnetate Tc 99m injection.
Rx only
See package insert for directions for use.
Tc-99m Half Life = 6 hr
CAUTION
RADIOACTIVE MATERIAL
Patent/Brevet/Patente:
www.curiumpharma.com/us-patents
Manufactured by/Fabrique par/Fabricado por:
Curium US LLC
Maryland Heights, MO 63043
Made in USA/Fabrique aux Etats-Unis/Hecho en E.U.A.
Distributed in Canada by/
Distribue au Canda par:
Curium Canada Inc.
Laval, QC, Canada, H7T-2R3
DIN: 02441608
CURIUM TM/MC
A901C0
R03/2019
Label
Ultra-Technekow™ V4
(TECHNETIUM Tc 99m GENERATOR)
Parent Molybdenum-99 prepared from Fission Produced Molybdenum-99
Store at Controlled Room Temperature 20-25ºC (68-77ºF)
WARNING: EACH UTK-V4 GENERATOR ELUATE MUST BE TESTED FOR MOLYBDENUM-99 CONTENT BEFORE USE.
Eluate must be used within 12 hours of elution.
For the production of sterile, non-pyrogenic Sodium Pertechnetate Tc 99m injection.
Rx only
See package insert for directions for use.
Tc-99m Half Life = 6 hr
CAUTION
RADIOACTIVE MATERIAL
Patent/Brevet/Patente:
www.curiumpharma.com/us-patents
Manufactured by/Fabrique par/Fabricado por:
Curium US LLC
Maryland Heights, MO 63043
Made in USA/Fabrique aux Etats-Unis/Hecho en E.U.A.
Distributed in Canada by/
Distribue au Canda par:
Curium Canada Inc.
Laval, QC, Canada, H7T-2R3
DIN: 02441608
CURIUM TM/MC
A901C0
R03/2019
Label
Ultra-Technekow™ V4
(TECHNETIUM Tc 99m GENERATOR)
Parent Molybdenum-99 prepared from Fission Produced Molybdenum-99
Store at Controlled Room Temperature 20-25ºC (68-77ºF)
WARNING: EACH UTK-V4 GENERATOR ELUATE MUST BE TESTED FOR MOLYBDENUM-99 CONTENT BEFORE USE.
Eluate must be used within 12 hours of elution.
For the production of sterile, non-pyrogenic Sodium Pertechnetate Tc 99m injection.
Rx only
See package insert for directions for use.
Tc-99m Half Life = 6 hr
CAUTION
RADIOACTIVE MATERIAL
Patent/Brevet/Patente:
www.curiumpharma.com/us-patents
Manufactured by/Fabrique par/Fabricado por:
Curium US LLC
Maryland Heights, MO 63043
Made in USA/Fabrique aux Etats-Unis/Hecho en E.U.A.
Distributed in Canada by/
Distribue au Canda par:
Curium Canada Inc.
Laval, QC, Canada, H7T-2R3
DIN: 02441608
CURIUM TM/MC
A901C0
R03/2019
Label
Ultra-Technekow™ V4
(TECHNETIUM Tc 99m GENERATOR)
Parent Molybdenum-99 prepared from Fission Produced Molybdenum-99
Store at Controlled Room Temperature 20-25ºC (68-77ºF)
WARNING: EACH UTK-V4 GENERATOR ELUATE MUST BE TESTED FOR MOLYBDENUM-99 CONTENT BEFORE USE.
Eluate must be used within 12 hours of elution.
For the production of sterile, non-pyrogenic Sodium Pertechnetate Tc 99m injection.
Rx only
See package insert for directions for use.
Tc-99m Half Life = 6 hr
CAUTION
RADIOACTIVE MATERIAL
Patent/Brevet/Patente:
www.curiumpharma.com/us-patents
Manufactured by/Fabrique par/Fabricado por:
Curium US LLC
Maryland Heights, MO 63043
Made in USA/Fabrique aux Etats-Unis/Hecho en E.U.A.
Distributed in Canada by/
Distribue au Canda par:
Curium Canada Inc.
Laval, QC, Canada, H7T-2R3
DIN: 02441608
CURIUM TM/MC
A901C0
R03/2019
Label
Ultra-Technekow™ V4
(TECHNETIUM Tc 99m GENERATOR)
Parent Molybdenum-99 prepared from Fission Produced Molybdenum-99
Store at Controlled Room Temperature 20-25ºC (68-77ºF)
WARNING: EACH UTK-V4 GENERATOR ELUATE MUST BE TESTED FOR MOLYBDENUM-99 CONTENT BEFORE USE.
Eluate must be used within 12 hours of elution.
For the production of sterile, non-pyrogenic Sodium Pertechnetate Tc 99m injection.
Rx only
See package insert for directions for use.
Tc-99m Half Life = 6 hr
CAUTION
RADIOACTIVE MATERIAL
Patent/Brevet/Patente:
www.curiumpharma.com/us-patents
Manufactured by/Fabrique par/Fabricado por:
Curium US LLC
Maryland Heights, MO 63043
Made in USA/Fabrique aux Etats-Unis/Hecho en E.U.A.
Distributed in Canada by/
Distribue au Canda par:
Curium Canada Inc.
Laval, QC, Canada, H7T-2R3
DIN: 02441608
CURIUM TM/MC
A901C0
R03/2019
Label
Ultra-Technekow™ V4
(TECHNETIUM Tc 99m GENERATOR)
Parent Molybdenum-99 prepared from Fission Produced Molybdenum-99
Store at Controlled Room Temperature 20-25ºC (68-77ºF)
WARNING: EACH UTK-V4 GENERATOR ELUATE MUST BE TESTED FOR MOLYBDENUM-99 CONTENT BEFORE USE.
Eluate must be used within 12 hours of elution.
For the production of sterile, non-pyrogenic Sodium Pertechnetate Tc 99m injection.
Rx only
See package insert for directions for use.
Tc-99m Half Life = 6 hr
CAUTION
RADIOACTIVE MATERIAL
Patent/Brevet/Patente:
www.curiumpharma.com/us-patents
Manufactured by/Fabrique par/Fabricado por:
Curium US LLC
Maryland Heights, MO 63043
Made in USA/Fabrique aux Etats-Unis/Hecho en E.U.A.
Distributed in Canada by/
Distribue au Canda par:
Curium Canada Inc.
Laval, QC, Canada, H7T-2R3
DIN: 02441608
CURIUM TM/MC
A901C0
R03/2019
Label
Ultra-Technekow™ V4
(TECHNETIUM Tc 99m GENERATOR)
Parent Molybdenum-99 prepared from Fission Produced Molybdenum-99
Store at Controlled Room Temperature 20-25ºC (68-77ºF)
WARNING: EACH UTK-V4 GENERATOR ELUATE MUST BE TESTED FOR MOLYBDENUM-99 CONTENT BEFORE USE.
Eluate must be used within 12 hours of elution.
For the production of sterile, non-pyrogenic Sodium Pertechnetate Tc 99m injection.
Rx only
See package insert for directions for use.
Tc-99m Half Life = 6 hr
CAUTION
RADIOACTIVE MATERIAL
Patent/Brevet/Patente:
www.curiumpharma.com/us-patents
Manufactured by/Fabrique par/Fabricado por:
Curium US LLC
Maryland Heights, MO 63043
Made in USA/Fabrique aux Etats-Unis/Hecho en E.U.A.
Distributed in Canada by/
Distribue au Canda par:
Curium Canada Inc.
Laval, QC, Canada, H7T-2R3
DIN: 02441608
CURIUM TM/MC
A901C0
R03/2019
Label
Ultra-Technekow™ V4
(TECHNETIUM Tc 99m GENERATOR)
Parent Molybdenum-99 prepared from Fission Produced Molybdenum-99
Store at Controlled Room Temperature 20-25ºC (68-77ºF)
WARNING: EACH UTK-V4 GENERATOR ELUATE MUST BE TESTED FOR MOLYBDENUM-99 CONTENT BEFORE USE.
Eluate must be used within 12 hours of elution.
For the production of sterile, non-pyrogenic Sodium Pertechnetate Tc 99m injection.
Rx only
See package insert for directions for use.
Tc-99m Half Life = 6 hr
CAUTION
RADIOACTIVE MATERIAL
Patent/Brevet/Patente:
www.curiumpharma.com/us-patents
Manufactured by/Fabrique par/Fabricado por:
Curium US LLC
Maryland Heights, MO 63043
Made in USA/Fabrique aux Etats-Unis/Hecho en E.U.A.
Distributed in Canada by/
Distribue au Canda par:
Curium Canada Inc.
Laval, QC, Canada, H7T-2R3
DIN: 02441608
CURIUM TM/MC
A901C0
R03/2019
Label
Ultra-Technekow™ V4
(TECHNETIUM Tc 99m GENERATOR)
Parent Molybdenum-99 prepared from Fission Produced Molybdenum-99
Store at Controlled Room Temperature 20-25ºC (68-77ºF)
WARNING: EACH UTK-V4 GENERATOR ELUATE MUST BE TESTED FOR MOLYBDENUM-99 CONTENT BEFORE USE.
Eluate must be used within 12 hours of elution.
For the production of sterile, non-pyrogenic Sodium Pertechnetate Tc 99m injection.
Rx only
See package insert for directions for use.
Tc-99m Half Life = 6 hr
CAUTION
RADIOACTIVE MATERIAL
Patent/Brevet/Patente:
www.curiumpharma.com/us-patents
Manufactured by/Fabrique par/Fabricado por:
Curium US LLC
Maryland Heights, MO 63043
Made in USA/Fabrique aux Etats-Unis/Hecho en E.U.A.
Distributed in Canada by/
Distribue au Canda par:
Curium Canada Inc.
Laval, QC, Canada, H7T-2R3
DIN: 02441608
CURIUM TM/MC
A901C0
R03/2019
Label
Ultra-Technekow™ V4
(TECHNETIUM Tc 99m GENERATOR)
Parent Molybdenum-99 prepared from Fission Produced Molybdenum-99
Store at Controlled Room Temperature 20-25ºC (68-77ºF)
WARNING: EACH UTK-V4 GENERATOR ELUATE MUST BE TESTED FOR MOLYBDENUM-99 CONTENT BEFORE USE.
Eluate must be used within 12 hours of elution.
For the production of sterile, non-pyrogenic Sodium Pertechnetate Tc 99m injection.
Rx only
See package insert for directions for use.
Tc-99m Half Life = 6 hr
CAUTION
RADIOACTIVE MATERIAL
Patent/Brevet/Patente:
www.curiumpharma.com/us-patents
Manufactured by/Fabrique par/Fabricado por:
Curium US LLC
Maryland Heights, MO 63043
Made in USA/Fabrique aux Etats-Unis/Hecho en E.U.A.
Distributed in Canada by/
Distribue au Canda par:
Curium Canada Inc.
Laval, QC, Canada, H7T-2R3
DIN: 02441608
CURIUM TM/MC
A901C0
R03/2019
Label
ULTRA-TECHNEKOW
V4
technetium tc-99m injection, solution |
|||||||||||||||
|
|||||||||||||||
|
|||||||||||||||
|
|||||||||||||||
|
ULTRA-TECHNEKOW
V4
technetium tc-99m injection, solution |
|||||||||||||||
|
|||||||||||||||
|
|||||||||||||||
|
|||||||||||||||
|
ULTRA-TECHNEKOW
V4
technetium tc-99m injection, solution |
|||||||||||||||
|
|||||||||||||||
|
|||||||||||||||
|
|||||||||||||||
|
ULTRA-TECHNEKOW
V4
technetium tc-99m injection, solution |
|||||||||||||||
|
|||||||||||||||
|
|||||||||||||||
|
|||||||||||||||
|
ULTRA-TECHNEKOW
V4
technetium tc-99m injection, solution |
|||||||||||||||
|
|||||||||||||||
|
|||||||||||||||
|
|||||||||||||||
|
ULTRA-TECHNEKOW
V4
technetium tc-99m injection, solution |
|||||||||||||||
|
|||||||||||||||
|
|||||||||||||||
|
|||||||||||||||
|
ULTRA-TECHNEKOW
V4
technetium tc-99m injection, solution |
|||||||||||||||
|
|||||||||||||||
|
|||||||||||||||
|
|||||||||||||||
|
ULTRA-TECHNEKOW
V4
technetium tc-99m injection, solution |
|||||||||||||||
|
|||||||||||||||
|
|||||||||||||||
|
|||||||||||||||
|
ULTRA-TECHNEKOW
V4
technetium tc-99m injection, solution |
|||||||||||||||
|
|||||||||||||||
|
|||||||||||||||
|
|||||||||||||||
|
ULTRA-TECHNEKOW
V4
technetium tc-99m injection, solution |
|||||||||||||||
|
|||||||||||||||
|
|||||||||||||||
|
|||||||||||||||
|
ULTRA-TECHNEKOW
V4
technetium tc-99m injection, solution |
|||||||||||||||
|
|||||||||||||||
|
|||||||||||||||
|
|||||||||||||||
|
ULTRA-TECHNEKOW
V4
technetium tc-99m injection, solution |
|||||||||||||||
|
|||||||||||||||
|
|||||||||||||||
|
|||||||||||||||
|
ULTRA-TECHNEKOW
V4
technetium tc-99m injection, solution |
|||||||||||||||
|
|||||||||||||||
|
|||||||||||||||
|
|||||||||||||||
|
Labeler - Curium US LLC (079875617) |
Mark Image Registration | Serial | Company Trademark Application Date |
---|---|
ULTRA-TECHNEKOW 72361585 0927758 Live/Registered |
MALLINCKRODT CHEMICAL WORKS 1970-06-03 |