PMA P880086S156

Device
AFFINITY/INTEGRITY/VICTORY FAMILY OF PACEMAKERS
Applicant
ABBOTT MEDICAL
PMA number
P880086
Supplement
S156
Product code
NVZ
Decision date
2007-12-12
Classification
Pulse Generator, Permanent, Implantable
Generic name
Pulse generator, permanent, implantable
Approval order statement
APPROVAL FOR TWO CORRECTIVE FIXES TO ADDRESS A HARDWARE ISSUED THAT CAN LEAVE THE CRT-D AND ICD SYSTEMS LISTED ABOVE IN A POST-SENSE REFRACTORY STATE. FOR DEVICES THAT ARE ALREADY IMPLANTED AND/OR ALREADY MANUFACTURED, CHANGES WILL BE MADE TO THE PROGRAMMER SOFTWARE. DURING A PATIENT FOLLOW-UP VISIT, THE PROGRAMMER WILL INTERROGATE THE DEVICE AND DOWNLOAD FIRMWARE RAM CODE TO CORRECT THE ANOMALOUS CONDITION IF IT OCCURS. FOR NEW DEVICES, A MODIFICATION HAS BEEN MADE TO THE SIO2 CHIP HARDWARE TO CHECK THE SENSE REFRACTORY STATE EVERY TWO SECONDS AND END IT IF APPROPRIATE.

Current openFDA PMA Record#

Device
AFFINITY/INTEGRITY/VICTORY FAMILY OF PACEMAKERS
Applicant
ABBOTT MEDICAL
PMA number
P880086
Supplement
S156
Product code
NVZ
Generic name
Pulse generator, permanent, implantable
Decision date
2007-12-12
Decision code
APPR
Date received
2007-10-24
Supplement type
Real-Time Process
Supplement reason
Change Design/Components/Specifications/Material
Approval order statement
APPROVAL FOR TWO CORRECTIVE FIXES TO ADDRESS A HARDWARE ISSUED THAT CAN LEAVE THE CRT-D AND ICD SYSTEMS LISTED ABOVE IN A POST-SENSE REFRACTORY STATE. FOR DEVICES THAT ARE ALREADY IMPLANTED AND/OR ALREADY MANUFACTURED, CHANGES WILL BE MADE TO THE PROGRAMMER SOFTWARE. DURING A PATIENT FOLLOW-UP VISIT, THE PROGRAMMER WILL INTERROGATE THE DEVICE AND DOWNLOAD FIRMWARE RAM CODE TO CORRECT THE ANOMALOUS CONDITION IF IT OCCURS. FOR NEW DEVICES, A MODIFICATION HAS BEEN MADE TO THE SIO2 CHIP HARDWARE TO CHECK THE SENSE REFRACTORY STATE EVERY TWO SECONDS AND END IT IF APPROPRIATE.