Recall Z-0327-2027

Recall number
Z-0327-2027
Event number
100070
Firm
GE Medical Systems, LLC
Firm FEI
2126677
Product code
MUE
Status
Open, Classified
Initiated
2026-08-28
Posted
2026-11-04
Root cause
Radiation Control for Health and Safety Act
510(k) numbers
K162268, K193166, K130704, K190809, K163252, K254090, K210151, K202902, K163302, K133972, K111361, K193334, K110729, K110798, K123520, K151645, K211215, K121674, K260087, K131885, K211720, K202822, K170930, K120246, K173576, K123995, K110572, K220073, K103485, K110717

Product#

Senographe Pristina

Reason for Recall#

GE HealthCare has become aware of a nonconformance to certain Senographe Pristina VIA systems. Specifically, in cases of 3D+2D manual mode where the acquisition configuration of the anode track/filter combination is different, it can result in the following: the collimator blades might be visible in the image, or the collimator might be open more than expected, and an incorrect collimator filter might be used with respect to the anode track selected This issue could result in a minor additional X-ray dose being delivered to the patient and obvious artifacts in the image (collimator blades).