Recall Z-0347-2027

Recall number
Z-0347-2027
Event number
100044
Firm
SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION
Firm FEI
3002808139
Product code
OWB
Status
Open, Classified
Initiated
2025-09-30
Posted
2026-11-03
Root cause
Process control
510(k) numbers
K221922, K203535, K231692, K252503, K250660, K232344, K113027, K061534, K121772, K130893, K243735, K210900, K233107, K173639, K050489, K252068, K222625, K232420, K161563, K243411, K212383, K190497, K190768, K202360, K081073, K181767, K170130, K251602, K153232, K213768

Product#

Trinias Opera Smart (C16s, F12s)/Opera Grand(C12s,C16s, F12s)/Smart OBL Interventional Fluoroscopic X-Ray System containing C-arm support (MH-500, MH-60),

Reason for Recall#

The mounting bolts securing the C-arm drive unit of the X-ray system may not be torqued to specification, potentially compromising structural stability and safe operation.